Measuring guide
1. Choose the frequency weighting
- A (dBA): models the sensitivity of human hearing, used for environmental and occupational noise.
- C (dBC): keeps low frequencies almost unchanged, used for music, machinery and LCpeak.
- Z (dBZ): no weighting, for technical measurements and spectrum analysis.
2. Choose the time weighting
- Fast (F, 125 ms): rapidly changing noise, the default for most measurements.
- Slow (S, 1 s): steady noise; the exposure dose is computed on this level.
- Impulse (I): hammering, short bangs.
3. Calibration
Microphones differ by up to 10 dB between devices. Before relying on a reading for anything important:
- With an acoustic calibrator (94 dB or 114 dB at 1 kHz): pick the matching method in Settings, measure for 10 seconds, apply.
- With a reference sound level meter: place both side by side in front of a steady noise source, measure for 10 seconds at the same time, enter the reference LAeq.
- If you already know the device offset: enter it manually.
Calibration profiles are stored on the device (synced with TsuDEV Pro), but they are only valid for that exact device.
4. Reading the results
- Leq: the average energy level of the whole session, the number to compare against regulations.
- LAF10 / LAF90: the level exceeded 10% / 90% of the time; LAF90 is the background noise.
- Dose: 100% is the exposure limit for an 8-hour working day.
5. Limits to keep in mind
A phone is not a certified class 1/class 2 sound level meter. Phone microphones usually saturate around 100-120 dB and lose sensitivity below 50 Hz. Use Sound Db to survey, monitor and compare; for legally binding measurements, use certified equipment.